Divyang Patel1, Apichaya Sripariwuth2, Mostafa Abozeed2
1Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio.
Objectives:
This study sought to retrospectively investigate outcomes of lead extraction by using pre-operative computed tomography (CT) scans to identify risk of complicated lead extraction to aid in pre-procedural planning.
Background:
Transvenous lead extractions remain high-risk procedures requiring specialized operators, equipment, and surgical backup. Data are lacking for how to identify difficult lead extractions. CT scans, which can illustrate the proximity of the lead to adherent venous structures can potentially aid in identifying difficult lead extractions.
Methods:
All cases of patients who were undergoing transvenous lead extractions at the authors' institution between 2015 and 2018, who had a pre-operative CT scan prior to lead extraction, were reviewed. The images were retrospectively reviewed to examine adherence of leads to the surrounding vein and obtained procedural outcomes.
Results:
A total of 203 cases were reviewed of patients undergoing transvenous lead extraction who had a pre-operative CT scan, and scans were separated based on lead location in the superior vena cava, as assessed by CT imaging. Scans were divided into 3 groups: those in a central location or <1 cm adherence (n = 28); those that had at least 1 lead with tip adherent >1 cm (n = 137); or those that had at least 1 lead outside the vein contour (n = 38). Although there was only 1 serious complication requiring vascular surgery intervention, patients with at least 1 lead outside the vein contour required significantly longer procedural time (190.8 ± 86.6 min vs. 158.1 ± 73.7 min vs. 142.8 ± 52.2 min; p = 0.019) and fluoroscopy time (33.1 ± 24.2 min vs. 19.6 ± 18.4 min vs. 18.3 ± 16.4 min; p = 0.0006) than those with leads adhering >1 cm and centrally located leads, respectively.
Conclusions:
Pre-operative CT scanning can identify difficult lead extractions prior to performing the procedure. This information may aid electrophysiologists in the planning of extraction procedures. Future prospective studies are needed to confirm these findings.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization