Related Experiment Video
Updated: Jan 1, 2026

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Further Evolution of a New Nonbiological Transcatheter Valvular Prosthesis
Filip Schröter1, Martin Hartrumpf1, Ralf-Uwe Kuehnel1
1Department of Cardiovascular Surgery, Heart Center Brandenburg, Brandenburg Medical School, Bernau bei Berlin, Brandenburg, Germany.
Background:
Polymeric heart valves are constructed from flexible synthetic materials, therefore aiming to combine the advantageous hemodynamic of biological and the longevity of mechanical valve prostheses. One such valve prototype in development is the PIZZA valve constructed of flexible triangular silicone leaflets on a foldable metal base for perspective transcatheter implantation. Here we present further improvements in its performance through structural modifications.
Methods:
Structurally modified prototypes were constructed from silicone sheets and stainless-steel wires. Their performance was then tested in a hemodynamic testing device of the type HKP 2.0.
Results:
Shift from a planar to a cone shape as well as overlapping of the leaflets significantly improved the valves performance, reducing regurgitation as well as systolic pressure gradients.
Conclusions:
The results of the modified prototypes expressed superior performance and represented a step forward on the road to an easily producible, polymeric transcatheter valvular prosthesis.