Comparison of Operator Workloads Associated with the Single-unit Anyfusion® Pump and the Changeover from a Syringe
Jeong Jin Min1, Duk Kyung Kim2, Kwan Young Hong1
1Department of Anesthesiology and Pain Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Background:
A changeover from a syringe pump to an infusion pump may transiently interrupt drug administration and result in medication errors. A newly developed infusion device (Anyfusion®) that combines the functions of an infusion pump and a syringe pump in a single unit was recently commercialized. This study compared the operator workloads associated with the Anyfusion® pump and with the conventional method using two infusion devices (a syringe pump and an infusion pump).
Methods:
This was a prospective, randomized, crossover-designed simulation study using an imitation arm with an intravascular line. We compared the two methods in terms of total execution time, total number of button clicks, and subjective difficulty using a numerical rating scale with a score from 0 (extremely easy) to 10 (extremely difficult).
Results:
Twenty-two nurses successfully performed both interventions according to the allocated sequence. Total execution times did not differ between the two methods (129.5 ± 23.2 seconds for the conventional method vs. 121.2 ± 24.3 seconds for the Anyfusion® method; P = 0.244), although the total number of clicks was significantly fewer using the Anyfusion® than the conventional method (median [interquartile range]: 10.0 [9.0-12.0] vs. 21.0 [20.0-25.0], respectively; P < 0.001). Participants rated the Anyfusion® method as easier than the conventional method (1.7 ± 1.2 vs. 3.6 ± 1.6; P < 0.001, respectively).
Conclusion:
The introduction of Anyfusion® lessened the workload of practitioners required by the changeover process, which might reduce the risk of medication errors and subsequent patient harm.
Trial Registration:
Clinical Research Information Service Identifier: KCT0004172.
More Related Videos
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Compartment Models: Single-Compartment Model
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...