Supercritical fluid chromatography-mass spectrometry in routine anti-doping analyses: Estimation of retention time
Gioacchino Luca Losacco1, Elisa Marconetto1, Raul Nicoli2
1Institute of Pharmaceutical Sciences of Western Switzerland, University of Geneva, CMU-Rue Michel Servet 1, 1211 Geneva 4, Switzerland.
Abstract:
The aim of this study was to estimate the retention time variability under reproducible conditions of an SFC-MS analytical method for routine anti-doping analyses. For this purpose, a set of 51 doping agents, as neat standards and spiked in diluted urine, was used to assess their retention times variability over a period of four months, as well as the column inter-batch reproducibility. Three UHPSFC stationary phases have been employed, the Acquity UPC2 Torus 2-Picolylamine (2-PIC), UPC2 Viridis BEH and Acquity UPLC HSS C18 SB. Four columns, per column chemistry, have been purchased to represent three different production lots, with a total of twelve columns employed in this study. The two columns from the same lot were applied to the first part of the study (repeatability), whereas the representative of three different lots were employed in the second part (robustness). In terms of organic modifier, a mixture of 98% MeOH and 2% water containing 20 mM ammonium formate was selected in order to limit the formation of methyl-silyl ethers on the surface of the silica particles, thus potentially improving the repeatability of retention times. A comparison with an UHPLC reference analytical method was made with the same set of analytes. The average relative standard deviations (RSD%), represented in split violin plots, illustrate how two of the UHPSFC columns assessed in this study were able to generate an excellent repeatability of retention times, with results that are in a similar range of those generated by UHPLC. Moreover, the Torus 2-PIC has proven to be the best of the three stationary phases, with an impressive RSD% of 0.5% in diluted urine relative to the inter-month variability. Finally, the inter-batch reproducibility assessment has highlighted a good reproducibility of the same stationary phase belonging to different production lots for all three column chemistries assessed, with the Viridis BEH silica generating an RSD% of 0.7% in diluted urine. Higher values of RSD (%) were found for Torus 2-PIC and HSS C18 SB, respectively of 1.0% and 1.6%.
More Related Videos
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: