A Magnetic Resonance Imaging Protocol for Stroke Onset Time Estimation in Permanent Cerebral Ischemia
Published on: September 16, 2017
14.5K
Phase-based T2 mapping with gradient echo imaging
Xiaoke Wang1,2, Diego Hernando1,3, Scott B Reeder1,2,3,4,5
1Department of Radiology, University of Wisconsin, Madison, Wisconsin.
Magnetic Resonance in Medicine
|December 25, 2019
Summary
This study introduces a novel phase-based MRI technique for faster transverse relaxation time (T2) mapping. The method accurately measures T2 in phantoms and in vivo, offering a promising alternative to conventional approaches.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Biomedical Engineering
- Medical Physics
Background:
- Transverse relaxation time (T2) mapping is crucial for MRI applications.
- Existing T2 mapping methods often use spin-echo (SE) sequences, which are time-consuming.
- There is a need for faster and efficient T2 mapping techniques.
Purpose of the Study:
- To develop and validate a novel phase-based T2 mapping technique using rapid gradient echo (GRE) acquisitions.
- To encode T2 information into the signal phase for improved mapping efficiency.
- To demonstrate the feasibility and accuracy of the proposed method.
Main Methods:
- Utilized Bloch equation simulations to establish the T2-dependent phase behavior in GRE acquisitions.
- Acquired at least two datasets with different RF phase increments to isolate T2-dependent phase.
- Validated the technique in phantom experiments and in vivo across various anatomical regions (brain, knee, abdomen, pelvis).
Main Results:
- Phase-based T2 maps showed accurate results compared to SE-based T2 mapping in phantoms.
- Good qualitative agreement was observed between the proposed method and reference techniques in vivo.
- In vivo T2 measurements in different anatomies correlated well with literature values.
Conclusions:
- A novel phase-based T2 mapping technique has been successfully developed.
- The technique demonstrates feasibility and accuracy in both phantom and in vivo studies.
- This method offers a potentially faster and efficient alternative for T2 quantification in MRI.