Related Experiment Video
Updated: Dec 31, 2025

Author Spotlight: Exploring Venous Waveforms for Non-Invasive Respiratory Monitoring in Pigs
Published on: March 8, 2024
Variability in the use of pulse oximeters with children in Kenyan hospitals: A mixed-methods analysis
Abigail J Enoch1, Mike English2,3,
1Nuffield Department of Population Health, University of Oxford, Oxford, United Kingdom (former DPhil student).
Insights
Pulse oximetry use in Kenyan hospitals is limited by supply, broken devices, and inadequate training. Improving availability and training can enhance its impact on pediatric care and oxygen therapy targeting.
Area of Science:
- Global Health
- Medical Technology Implementation
- Pediatric Critical Care
Background:
- Pulse oximetry is a cost-effective technology with potential to improve pediatric health outcomes.
- Widespread adoption of pulse oximetry in low- and middle-income countries remains a challenge.
- Barriers to pulse oximetry use in Kenyan hospitals require investigation.
Purpose of the Study:
- To determine the patterns of pulse oximeter use in Kenyan hospitals for children.
- To assess the impact of pulse oximetry on treatment decisions, particularly oxygen provision.
- To identify barriers hindering the effective implementation and utilization of pulse oximetry.
Main Methods:
- Analysis of 27,906 child admissions data from Kenya's Clinical Information Network (CIN) (2013-2016).
- Multivariable regression models and qualitative interviews with 30 healthcare workers across 14 hospitals.
- Adaptation of the Integrative Model of Behavioural Prediction to link quantitative and qualitative findings.
Main Results:
- Half of admitted children received a pulse oximeter reading; 10% had oxygen saturation below 90%.
- Pulse oximetry use was associated with increased likelihood of oxygen prescription, especially for saturation <90%.
- Key barriers included inadequate supply, malfunctioning devices, and insufficient training on use and interpretation.
Conclusions:
- Significant challenges persist in implementing pulse oximetry into routine Kenyan pediatric care.
- Recommendations include improving supply chain management, repair systems, training, and clinical documentation.
- Effective implementation is crucial for realizing the benefits of pulse oximetry in guiding oxygen therapy and improving patient outcomes.
Background:
Pulse oximetry, a relatively inexpensive technology, has the potential to improve health outcomes by reducing incorrect diagnoses and supporting appropriate treatment decisions. There is evidence that in low- and middle-income countries, even when available, widespread uptake of pulse oximeters has not occurred, and little research has examined why. We sought to determine when and with which children pulse oximeters are used in Kenyan hospitals, how pulse oximeter use impacts treatment provision, and the barriers to pulse oximeter use.
Methods And Findings:
We analyzed admissions data recorded through Kenya's Clinical Information Network (CIN) between September 2013 and February 2016. We carried out multiple imputation and generated multivariable regression models in R. We also conducted interviews with 30 healthcare workers and staff from 14 Kenyan hospitals to examine pulse oximetry adoption. We adapted the Integrative Model of Behavioural Prediction to link the results from the multivariable regression analyses to the qualitative findings. We included 27,906 child admissions from 7 hospitals in the quantitative analyses. The median age of the children was 1 year, and 55% were male. Three-quarters had a fever, over half had a cough; other symptoms/signs were difficulty breathing (34%), difficulty feeding (34%), and indrawing (32%). The most common diagnoses were pneumonia, diarrhea, and malaria: 45%, 35%, and 28% of children, respectively, had these diagnoses. Half of the children obtained a pulse oximeter reading, and of these, 10% had an oxygen saturation level below 90%. Children were more likely to receive a pulse oximeter reading if they were not alert (odds ratio [OR]: 1.30, 95% confidence interval (CI): 1.09, 1.55, p = 0.003), had chest indrawing (OR: 1.28, 95% CI: 1.17, 1.40, p < 0.001), or a very high respiratory rate (OR: 1.27, 95% CI: 1.13, 1.43, p < 0.001), as were children admitted to certain hospitals, at later time periods, and when a Paediatric Admission Record (PAR) was used (OR PAR used compared with PAR not present: 2.41, 95% CI: 1.98, 2.94, p < 0.001). Children were more likely to be prescribed oxygen if a pulse oximeter reading was obtained (OR: 1.42, 95% CI:1.25, 1.62, p < 0.001) and if this reading was below 90% (OR: 3.29, 95% CI: 2.82, 3.84, p < 0.001). The interviews indicated that the main barriers to pulse oximeter use are inadequate supply, broken pulse oximeters, and insufficient training on how, when, and why to use pulse oximeters and interpret their results. According to the interviews, variation in pulse oximeter use between hospitals is because of differences in pulse oximeter availability and the leadership of senior doctors in advocating for pulse oximeter use, whereas variation within hospitals over time is due to repair delays. Pulse oximeter use increased over time, likely because of the CIN's feedback to hospitals. When pulse oximeters are used, they are sometimes used incorrectly and some healthcare workers lack confidence in readings that contradict clinical signs. The main limitations of the study are that children with high levels of missing data were not excluded, interview participants might not have been representative, and the interviews did not enable a detailed exploration of differences between counties or across senior management groups.
Conclusions:
There remain major challenges to implementing pulse oximetry-a cheap, decades old technology-into routine care in Kenya. Implementation requires efficient and transparent procurement and repair systems to ensure adequate availability. Periodic training, structured clinical records that include prompts, the promotion of pulse oximetry by senior doctors, and monitoring and feedback might also support pulse oximeter use. Our findings can inform strategies to support the use of pulse oximeters to guide prompt and effective treatment, in line with the Sustainable Development Goals. Without effective implementation, the potential benefits of pulse oximeters and possible hospital cost-savings by targeting oxygen therapy might not be realized.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: