Related Experiment Video
Updated: Jan 10, 2026

Optimization of Antigen Preparation for Hemagglutination Inhibition Assay of Newcastle Disease Virus Serology
Published on: January 24, 2025
Reduction of variation in results of rheumatoid factor tests by use of a serum reference preparation
Abstract:
Standaridizing test results for rheumatoid factor by comparing results obtained for an unknown with results obtained for a serum reference preparation decreased variance between laboratories, as measured in the Center for Disease Control proficiency testing program, by 77%. The amount of improvement was also estimated by the type of test and by the manufacturer's product. Standardization resulted in an increase in the number of reported results that were within a twofold dilution of the median value. The percentage increased from 50.3 to 93.7% for the slide tests and from 78.1 to 91.2% fro the tube tests. Decrease in variance by manufacturer's product ranged from 94 to 27%. The study demonstrated that adopting a reference serum standard could substantially improve the comparability of rheumatoid factor test results and that proficiency testing programs can be used to estimate improvement which could be expected as a result of standardization.