A comparison between EPSON V700 and EPSON V800 scanners for film dosimetry
Tarafder Shameem1,2, Nick Bennie3, Martin Butson4,5
1North Coast Cancer Institute, Lismore, NSW, Australia. tarafder.shameem@health.nsw.gov.au.
Abstract:
Radiochromic film is a good dosimeter choice for patient QA for complex treatment techniques (IMRT, VMAT, SABR, SBRT) because of its near tissue equivalency, very high spatial resolution and established method of use. Commercial scanners are usually used for film dosimetry, among which EPSON scanners are the most common. NCCI have used an EPSON V700 scanner, but recently acquired a new model EPSON V800 scanner. The purpose of this work was to evaluate any differences between these two scanners to consider whether they can be used interchangeably or not. Different aspects of film dosimetry, e.g. lateral response artefact (LRA) effect, orientation effect, scanner response etc., were compared. EBT3 films were irradiated with 40 × 40 cm2 field size 6 MV beams and scanned in both the scanners. The scanned images were read in ImageJ V1.49 software. The data obtained was then copied in MS Excel to compare the scanners. The V800 scanner causes more polarisation, which results in more LRA effect than for the V700 scanner. The responses of the scanners in all three colour channels are not the same for the same film and irradiation. The V800 scanner shows an increase of response of up to 1.6% compared to 3.7% increase in the V700 scanner after scanning a piece of irradiated film 20 times. The scanners cannot be used interchangeably. The correction factors for LRA effect and the calibration curves are different. Further characterisation, evaluation and commissioning is required before clinical use.
More Related Videos
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021