Design and evaluation of a laboratory-based wheelchair castor testing protocol using community data
Anand Mhatre1,2, Norman Reese2,3, Jon Pearlman1,2
1Department of Rehabilitation Science and Technology, University of Pittsburgh, Pittsburgh, Pennsylvania, United States of America.
Plos One
|January 11, 2020
Summary
Wheelchair castor failures are common, but current tests are insufficient. New testing including corrosion and abrasion better predicts real-world durability and identifies injury risks.
Area of Science:
- Biomedical Engineering
- Rehabilitation Engineering
- Materials Science
Background:
- Wheelchair castor failures lead to significant physical, social, and economic burdens for users.
- Existing wheelchair test methods and regulations may not adequately identify poorly designed castors prone to failure.
- The International Society of Wheelchair Professionals (ISWP) identified castor failure as a high priority, recommending a new testing system.
Purpose of the Study:
- To develop a new castor testing protocol incorporating community-relevant exposure data.
- To evaluate the impact of shock, corrosion, and abrasion on wheelchair castor durability.
- To assess how these combined exposures alter castor failure modes and associated injury risks.
Main Methods:
- A castor testing protocol was developed based on community shock, corrosion, and abrasion data.
- Eight different castor models were tested under four conditions: shock alone, corrosion + shock, abrasion + shock, and abrasion + corrosion + shock.
- A total of eight samples per model were evaluated across the four conditions.
Main Results:
- Corrosion and abrasion significantly reduced castor durability by 13% to 100% across tested models.
- The inclusion of corrosion and abrasion altered failure modes in 75% of tested castor models.
- Two-thirds of these altered failure modes were associated with an increased risk of injury to wheelchair users.
Conclusions:
- Community-based factors like corrosion and abrasion substantially decrease wheelchair castor durability.
- The developed testing protocol, including corrosion and abrasion, is crucial for accurately assessing castor performance.
- Incorporating these elements into wheelchair standards could improve user safety and reduce failures.