Related Experiment Video
Updated: Dec 31, 2025

Bidirectional Electrical and Optoelectronic Interfaces in Healthy and Ischemic Ex Vivo Rat Hearts
Published on: July 18, 2025
Long duration stabilization of porous silicon membranes in physiological media: Application for implantable reactors
Abdoullatif Baraket1, Jean-Pierre Alcaraz2, Chantal Gondran3
1ISA, Institut des Sciences Analytiques, Département LSA, 5, rue de la Doua, 69100, Villeurbanne, France.
Abstract:
The natural biodegradabilty of porous silicon (pSi) in physiological media limits its wider usage for implantable systems. We report the stabilization of porous silicon (pSi) membranes by chemical surface oxidation using RCA1 and RCA2 protocols, which was followed by a PEGylation process using a silane-PEG. These surface modifications stabilized the pSi to allow a long period of immersion in PBS, while leaving the pSi surface sufficiently hydrophilic for good filtration and diffusion of several biomolecules of different sizes without any blockage of the pSi structure. The pore sizes of the pSi membranes were between 5 and 20 nm, with the membrane thickness around 70 μm. The diffusion coefficient for fluorescein through the membrane was 2 × 10-10 cm2 s-1, and for glucose was 2.2 × 10-9 cm2 s-1. The pSi membrane maintained that level of glucose diffusion for one month of immersion in PBS. After 2 months immersion in PBS the pSi membrane continued to operate, but with a reduced glucose diffusion coefficient. The chemical stabilization of pSi membranes provided almost 1 week stable and functional biomolecule transport in blood plasma and opens the possibility for its short-term implantation as a diffusion membrane in biocompatible systems.