Methylene Diphosphonate Bone Scan Scintigraphic Image Enhancement using Gamma Correction and Optimizing the Value of
Anshul Sharma1, Anil Kumar Pandey1, Deepak Khichi1
1Department of Nuclear Medicine, All India Institute of Medical Science, New Delhi, India.
Introduction:
Focal areas of high radiotracer uptake in a bone-scan image can result in dynamic range of the intensity value to exceed the dynamic range of the display, requiring multiple interactive contrast adjustments. This unnecessary burden on time of physician can be avoided using power law equation to brighten up the low-intensity areas in image. However, despite the widespread availability of this technique in commercial systems, for this clinical setting, the gamma-value needs to be standardized.
Materials And Methods:
Sixty dark bone scan images were selected. Ten randomly selected images from this set were evaluated qualitatively and quantitatively (perception-based image quality evaluator, absolute mean brightness error, structural similarity, and peak signal to noise ratio) to select a range of gamma values (from 0.1 to 0.9, increment of 0.1), where the results were acceptable. This range of gamma was then applied to rest of the 50 images to find the best value. Images were evaluated by two experienced nuclear medicine physicians. Although not ideal, but for the purpose of simplicity, we tried reaching a single best value. For this, the physicians were asked to reach consensus on the acceptable images.
Results:
In the first part of the study, after evaluation of 100 images (1 original and 9 processed images with 0.1-0.9 gammas), range of gamma values from 0.3 to 0.8 was found to be optimum. This range was then applied to rest of the 50 images. Evaluation of resultant 350 images (1 original and 6 processed for each input image) further narrowed this range to 0.4-0.7 (0.3 gamma selected only twice by one physician). The kappa for acceptable images was moderate at 0.482 (P <0.001). The single gamma value of 0.6 resulted in 72% of the images to be acceptable.
Conclusion:
Use of power law equation to brighten up the low intensity areas of dark bone scan images, without loss of clinically significant details, is feasible with single gamma value of 0.6 and range of 0.4-0.7 giving best results.
More Related Videos
13:16Computed Tomography and Optical Imaging of Osteogenesis-angiogenesis Coupling to Assess Integration of Cranial Bone Autografts and Allografts
Published on: December 22, 2015
07:52In Vivo Imaging and Tracking of Technetium-99m Labeled Bone Marrow Mesenchymal Stem Cells in Equine Tendinopathy
Published on: December 9, 2015
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...