Pacemaker endless loop tachycardia: termination by simple techniques other than magnet application.

S S Barold1, M D Falkoff, L S Ong

  • 1Department of Medicine, Genesee, Rochester, New York 14607.

The American Journal of Medicine
|December 1, 1988
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Remaining agile - post-pandemic prevention of SARS-CoV-2 nosocomial transmission.

The Journal of hospital infection·2023
Same author

True vs. pseudo-electrical ventricular alternans sustained by ventricular premature complexes.

Herzschrittmachertherapie & Elektrophysiologie·2018
Same author

A review of the atrial upper rate algorithms of St. Jude Medical (Abbott) cardiac implantable electronic devices : Incidence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS).

Herzschrittmachertherapie & Elektrophysiologie·2017
Same author

Erratum to: Noise from a dysfunctional atrial lead detected as atrial fibrillation by a cardiac implantable electronic device.

Herzschrittmachertherapie & Elektrophysiologie·2017
Same author

Diagnosis of myocardial infarction and ischemia in the setting of bundle branch block and cardiac pacing.

Herzschrittmachertherapie & Elektrophysiologie·2016
Same author

Unusual cause of desynchronization in a cardiac resynchronization device.

Herzschrittmachertherapie & Elektrophysiologie·2012

Chest wall stimulation (CWS), myopotential oversensing, and chest thumping offer effective alternatives for terminating pacemaker endless loop tachycardia (ELT) when magnet use is not feasible. These simple methods provide accessible solutions for physicians managing pacemaker-dependent arrhythmias.

Area of Science:

  • Electrophysiology
  • Cardiovascular Medicine
  • Biomedical Engineering

Background:

  • Pacemaker endless loop tachycardia (ELT) is a common arrhythmia.
  • Magnet application is the standard method for terminating ELT by converting pacemakers to asynchronous mode.

Purpose of the Study:

  • To investigate alternative, simple methods for terminating ELT.
  • To evaluate chest wall stimulation (CWS), myopotential oversensing, and chest thumping as ELT termination strategies.

Main Methods:

  • A modified CWS technique using an external pulse generator was tested in 74 patients with DDD pacemakers.
  • CWS was applied during induced ELT in 20 patients.
  • Myopotential oversensing and chest thumping were attempted during ELT in subsets of patients.

Related Experiment Videos

Main Results:

  • Modified CWS successfully terminated ELT in all tested patients.
  • Myopotential oversensing terminated ELT in 60% of patients.
  • Chest thumping terminated ELT in 67% of patients.

Conclusions:

  • CWS, myopotential oversensing, and chest thumping are effective alternatives to magnet application for ELT termination.
  • These techniques are simple and can be readily applied by physicians, even those less familiar with advanced pacemaker programming.