Related Experiment Video
Updated: Dec 30, 2025

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
A Long-Term Complication Occurred After Transcatheter Closure of Large Atrial Septal Defect
Chenyu Zhao1, Jinyu Xu1, Dan Cui1
1Department of Cardiovascular Surgery, The Second Hospital of Jilin University.
Abstract:
Transcatheter closure of ostium secondum atrial septal defect has become an alternative method to surgical closure. However, the incidence of complications and long-term results of using large size (> 40 mm) Amplatzer septal occluders are unknown. This case reported a 59 years old woman, whom received transcatheter closure of atrial septal defect (36 mm) with a 40 mm Amplatzer septal occluder 10 years ago and was diagnosed with heart failure. Transthroacic echocardiography showed severe mitral valve regurgitation. Intra-operatively, we confirmed and removed the large device, but we found that the mitral annulus was badly damaged. Mitral valve replacement was performed. We believe large size devices need to be implanted cautiously, especially for the large defect with insufficient rims, and also routinely follow-up is necessary.
More Related Videos
Related Concept Videos
Aneurysm III: Interprofessional Care
Cardiac Catheterization II: Right Heart Catheterization
Rheumatic Heart Disease I: Introduction
Aortic Regurgitation I: Introduction
Mitral Stenosis I: Introduction
Aortic Regurgitation III: Medical Management