Related Experiment Video
Updated: Dec 30, 2025

Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Luminal Water Imaging: Comparison With Diffusion-Weighted Imaging (DWI) and PI-RADS for Characterization of Prostate
Stefanie J Hectors1,2,3, Daniela Said1,2,4, Jeffrey Gnerre2
1BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
Background:
Luminal water imaging (LWI), a multicomponent T2 mapping technique, has shown promise for prostate cancer (PCa) detection and characterization.
Purpose:
To 1) quantify LWI parameters and apparent diffusion coefficient (ADC) in PCa and benign peripheral zone (PZ) tissues; and 2) evaluate the diagnostic performance of LWI, ADC, and PI-RADS parameters for differentiation between low- and high-grade PCa lesions.
Study Type:
Prospective.
Subjects:
Twenty-six PCa patients undergoing prostatectomy (mean age 59 years, range 46-72 years).
Field Strength/Sequence:
Multiparametric MRI at 3.0T, including diffusion-weighted imaging (DWI) and LWI T2 mapping.
Assessment:
LWI parameters and ADC were quantified in index PCa lesions and benign PZ.
Statistical Tests:
Differences in MRI parameters between PCa and benign PZ were assessed using Wilcoxon signed tests. Spearman correlation of pathological grade group (GG) with LWI parameters, ADC, and PI-RADS was evaluated. The utility of each of the parameters for differentiation between low-grade (GG ≤2) and high-grade (GG ≥3) PCa was determined by Mann-Whitney U tests and ROC analyses.
Results:
Twenty-six index lesions were analyzed (mean size 1.7 ± 0.8 cm, GG: 1 [n = 1; 4%], 2 [n = 14, 54%], 3 [n = 8, 31%], 5 [n = 3, 12%]). LWI parameters and ADC both showed high diagnostic performance for differentiation between benign PZ and PCa (highest area under the curve [AUC] for LWI parameter T2,short [AUC = 0.98, P < 0.001]). The LWI parameters luminal water fraction (LWF) and amplitude of long T2 component Along significantly correlated with GG (r = -0.441, P = 0.024 and r = -0.414, P = 0.036, respectively), while PI-RADS, ADC, and the other LWI parameters did not (P = 0.132-0.869). LWF and Along also showed significant differences between low-grade and high-grade PCa (AUC = 0.776, P = 0.008 and AUC = 0.758, P = 0.027, respectively). Maximum diagnostic performance for discrimination of high-grade PCa was found with combined LWI parameters (AUC 0.891, P = 0.001).
Data Conclusion:
LWI parameters, in particular in combination, showed superior diagnostic performance for differentiation between low-grade and high-grade PCa compared to ADC and PI-RADS assessment. J. Magn. Reson. Imaging 2020;52:271-279.