Related Experiment Video
Updated: Dec 29, 2025

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Cup-cage construct for massive acetabular defect in revision hip arthroplasty- A case series with medium to long-term
Dan Arvinte1, Manish Kiran1, Manoj Sood1
1Department of Trauma and Orthopaedics, Bedford Hospital NHS Trust, Kempston Road, Bedford, MK42 9DJ, United Kingdom.
Objective:
Cup-cage reconstruction has emerged as a possible solution for managing massive acetabular defects with a few existing studies reporting encouraging results at mid-term follow-up. We present our experience with this unitised construct.
Method:
Six patients (7 hips) with a mean age of 76 years (73-81) were revised due to catastrophic aseptic failure of a primary cup implanted 10-19 years previously, having a Paprosky type 3B acetabular defect.
Results:
At a mean follow-up of 72 months (63-140) no cases have required re-revision. Oxford Hip Scores improved from an average of 8 (1-17) preoperatively to an average of 36 (18-45) at the last follow-up. WOMAC scores preoperatively averaged 76 (49-96) and postoperatively averaged 26.5 points (0-69) at the last follow-up. SF-12 scores improved in both components. One patient showed non-progressive osteolysis around the ischial flange and one had less than 5 mm migration of the construct. One patient died of unrelated causes.
Conclusion:
Our study presents one of the longest follow-up of cup-cage construct and supports the previously reported good results; it encourages the use of this construct in reconstruction of massive acetabular defect, with or without pelvic discontinuity.