Related Experiment Video
Updated: Dec 29, 2025

Movement Retraining using Real-time Feedback of Performance
Published on: January 17, 2013
Quality assurance of isocentres for passive proton beam nozzles using motion capture cameras
Wonjoong Cheon1, Kwanghyun Jo2, Sung Hwan Ahn2
1Department of Health Sciences and Technology, SAIHST, Sungkyunkwan University, Seoul 06351, Republic of Korea.
Purpose:
The objective of this work is to determine mechanical, radiation, and imaging isocentres in three-dimensional (3D) coordinates and verifying coincidence of isocentres of passively scattered proton beam using a visual tracking system (VTS) and an in-house developed phantom named the Eagle.
Methods:
The Eagle phantom consists of two modules: The first, named Eagle-head, is used for determining 3D mechanical isocentre of gantry rotation. The second, named Eagle-body, is used for determining 3D radiation and imaging isocentres. The Eagle-body has four slots wherein radiochromic films were inserted for measuring the 3D radiation isocentre and a metal bead was embedded in the centre of one cube to determine the imaging isocentre; this was determined by analysing cone-beam computed tomography images of the cube. Infrared reflective markers that can be tracked by VTS were attached to the Eagle at predetermined locations. The tracked data were converted into 3D treatment room coordinates. The developed method was compared with other methods to assess accuracy.
Results:
The isocentres were determined in mm with respect to the laser isocentre. The mechanical, radiation, and imaging isocentres were (-0.289, 0.189, 0.096), (-0.436, -0.217, 0.009), and (0.134, 0.142, 0.103), respectively. When compared with other methods, the difference in coordinates was (-0.033, -0.107, 0.014) and (0.003, 0.067, 0.039) for radiation and imaging isocentres, respectively.
Conclusion:
The developed system was found to be useful in providing fast and accurate measurements of the three isocentres in the 3D treatment room coordinate system.
More Related Videos
07:43In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
08:24Sit-to-stand-and-walk from 120% Knee Height: A Novel Approach to Assess Dynamic Postural Control Independent of Lead-limb
Published on: August 30, 2016