Related Experiment Video
Updated: Dec 29, 2025

Shotgun Lipidomics of Rodent Tissues
Published on: November 18, 2022
Interface for Reproducible, Multishot Direct Analysis of Solid-Phase Microextraction Samples
G Asher Newsome1, Gwénaëlle Kavich1, Alba Alvarez-Martin1
1Smithsonian Institution Museum Conservation Institute, 4210 Silver Hill Road, Suitland, Maryland 20746-2863, United States.
Abstract:
An enclosed interface that joins a direct analysis in real time (DART) probe, solid-phase microextraction (SPME) fiber, and the inlet of a high-resolution mass spectrometer is described. Unlike other systems to couple SPME sampling to ambient mass spectrometry, the interface is able to perform discrete analyses on different areas of a single SPME fiber device for up to three technical replicate measurements of one sampling event. Inlet flow speed and desorption temperature are optimized, and reproducibility is demonstrated between replicate analyses on the same derivatized SPME fiber and with sequential fiber sampling events, yielding analyte measurement center of variance (CV) from 3 to 6%. Conditioning is also performed with the enclosed DART. The interface is a straightforward addition to commercially available technologies, and machine diagrams for custom components operated with SPME/DART/MS equipment are included.

