Wearable Piezoelectric-Based System for Continuous Beat-to-Beat Blood Pressure Measurement
Ting-Wei Wang1, Shien-Fong Lin1,2,3
1Department of Electrical and Computer Engineering, College of Electrical and Computer Engineering, National Chiao Tung University, Hsinchu 30010, Taiwan.
Abstract:
Non-invasive continuous blood pressure measurement is an emerging issue that potentially can be applied to cardiovascular disease monitoring and prediction. Recently, many groups have proposed the pulse transition time (PTT) method to estimate blood pressure for long-term monitoring. However, the PTT-based methods for blood pressure estimation are limited by non-specific estimation models and require multiple calibrations. This study aims to develop a low-cost wearable piezoelectric-based system for continuous beat-to-beat blood pressure measurement. The pressure change in the radial artery was extracted by systolic and diastolic feature points in pressure pulse wave (PPW) and the pressure sensitivity of the sensor. The proposed system showed a reliable accuracy of systolic blood pressure (SBP) (mean absolute error (MAE) ± standard deviation (SD) 1.52 ± 0.30 mmHg) and diastolic blood pressure (DBP, MAE ± SD 1.83 ± 0.50), and its performance agreed with standard criteria of MAE within 5 mmHg and SD within ±8 mmHg. In conclusion, this study successfully developed a low-cost, high-accuracy piezoelectric-based system for continuous beat-to-beat SBP and DBP measurement without multiple calibrations and complex regression analysis. The system is potentially suitable for continuous, long-term blood pressure-monitoring applications.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.