Related Experiment Video
Updated: Dec 28, 2025

Short-Duration Hypothermia Induction in Rats using Models for Studies examining Clinical Relevance and Mechanisms
Published on: March 3, 2021
Impact of a Novel Hypothermia Alert Device on Death of Low Birthweight Babies at Four Weeks: A Non-randomized
Mona Sharma1, Virginia Morgan2, Murthy Siddadiah2
1BEMPU Health, Bengaluru, Karnataka, India. Correspondence to: Dr Mona Sharma, Head of Product Design and Customer Research, BEMPU Health, 3C Alsa Glenridge Apt, 32, Langford Rd, Shanti Nagar, Bengaluru, Karnataka 560 027, India. mona@bempu.com.
Objective:
To study the effectiveness of a hypothermia monitoring device in reducing neonatal mortality and increasing Kangaroo Mother Care compliance.
Design:
Non-randomized controlled trial.
Setting:
3 government Sick Newborn Care Units and 7 Primary Health Centers in Udaipur and Dungarpur districts of Rajasthan for 4 months. The follow-up period was 4 weeks for each baby.
Participants:
Total 386 neonates were included in the study. 250 (64.76%) new-borns in the study group (BEMPU bracelet) and 136 (35.23%) enrolled in the control group. Clinically stable babies discharged below 2500 grams, whose parents could be reached by phone, and who could visit the facility for 4 weekly follow-ups were eligible for participation. Infants with complications or those leaving against medical advice were not eligible.
Intervention:
The BEMPU Bracelet is a medical device that provides 4 weeks of continuous hypothermia monitoring for new-borns, and emits an audio-visual alarm when the temperature of the newborn is below 36.5°C.
Outcome:
Neonatal mortality over the 4-week period.
Results:
Mortality data was obtained for 92% (229 babies) of the study group and 91% of the control group (124 babies) at the end of the 4-week period. The intervention group had a significantly lower mortality rate as compared to the control group (6% vs. 14%, P=0.013). Weight data from 51% of the study group (128 babies) and 32% of the control group (44 babies) did not show a significant difference in weight gain between the groups.
Conclusions:
The observed effect on mortality and qualitative feedback on KMC compliance suggest the utility of the device in the community settings.