Validation of the Gel & Wax Boluses and Comparison of their Dosimetric Performance with Virtual Bolus
T R Verma1, N K Painuly1, M Tyagi2
1PhD, Department of Radiotherapy, King George's Medical University, UP, Lucknow, India.
Background:
In general, radiotherapy treatment planning is performed using the virtual bolus. It is necessary to investigate physical bolus in comparison to virtual one.
Objectives:
In the present study, first, radiological properties of superflab Gel bolus and Paraffin wax bolus was investigated in terms of their relative electron density. Then, dosimetric performance of both the bolus (i.e. Gel and Parafin wax) was compared with Virtual bolus.
Material And Methods:
In This experimental study, the radiological property of Wax and Gel boluses was investigated using two methods. In one, the relative electron density of both the Gel and Wax boluses was calculated by measuring their linear attenuation coefficient where in another method relative electron density was calculated by recording their CT No directly from their CT scan. Later CT scan of solid water slab phantom (dimension 30x30x15 cm3), with physical boluses (i.e. Gel and Wax bolus) of appropriate thicknesses required to deliver a dose of 200 cGy at Dmax using 4 MV, 6 MV and 15 MV photon beams, was taken. These CT data sets were retrieved to TPS. A plan was done to deliver a dose of 200 cGy at Dmax using Single 4 MV, 6 MV and 15 MV photon beams. Dose at depths Dmax, 1 cm, 2 cm, 3 cm, 4 cm and 5 cm was recorded. Using this similar method, doses at depths viz Dmax, 1 cm, 2 cm, 3 cm, 4 cm and 5 cm was recorded for the Gel and Wax boluses. The differences in dose of gel and wax bolus from virtual bolus were recorded for comparison of their dosimetric performance.
Results:
The measured (calculated) relative electron density of wax and Gel bolus was found to be 0.958 (0.926) and 0.923 (0.907), respectively. Variation in dosimetric performance of Gel and Wax with reference to Virtual bolus was studied. However, on average, Gel bolus was more consistent with virtual bolus.
Conclusion:
To avoid any dose difference between, delivered (using physical bolus) and planned (using virtual bolus), the physical boluses should be investigated for their dosimetric performance in comparison to virtual bolus. The results obtained and methodology used in this study can be applied in routine radiotherapy practices.
More Related Videos
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Bioequivalence studies: Biowaivers