Interlaboratory comparison of heavy metal testing in animal diagnostic specimens and feed using inductively coupled
Steven McGeehan1,2,3,4,5, Timothy Baszler1,2,3,4,5, Cynthia Gaskill1,2,3,4,5
1Washington Animal Disease Diagnostic Laboratory, College of Veterinary Medicine, Washington State University, Pullman, WA (Baszler, Talcott).
Abstract:
We compared inductively coupled plasma-mass spectrometry (ICP-MS) test results for the analysis of heavy metals (As, Ba, Cd, Hg, Pb, and Se) in pet foods and routine veterinary diagnostic specimens using intralaboratory and interlaboratory comparisons. Four laboratories, 1 principal laboratory and 3 collaborating laboratories, conducted instrument comparison (limit of detection [LOD], limit of quantification [LOQ], and linear dynamic range [LDR] on 24 data sets), in-house method comparison (accuracy and precision on 120 data sets), and interlaboratory comparison (reproducibility on 528 data sets using Horwitz equation analysis). Matrices tested included 2 types of pet food jerky treats (chicken and sweet potato), bovine blood, and bovine liver and kidney. The instrument comparison study confirmed that ICP-MS provided the sensitivity necessary for the analysis of all heavy metals tested at concentrations below the level of concern for routine diagnostic testing. The "in-house" method comparison samples, spiked at low (0.04 µg/g), medium (0.4 µg/g), and high (8.0 µg/g; note: the high validation level spike for mercury was 2 µg/g) concentration levels, indicated that ICP-MS can meet U.S. FDA acceptance criteria for both accuracy (90-105% recovery) and precision (< 6% coefficient of variation). The interlaboratory comparison studies showed that ICP-MS is a reproducible method for the analysis of heavy metals (HorRat value of 0.5-2.0) except for mercury in one laboratory, which used a different sample preparation method (open block rather than microwave digestion). Overall, our study showed that ICP-MS is a reproducible method for the analysis of heavy metals in spite of minor differences in methodology.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab