A systematic review of incubator-based neonatal radiography - What does the evidence say?

J Tugwell-Allsup1, A England2

  • 1Besti Cadwaladr University Health Board, Ysbyty Gwynedd, Pnerhosgarnedd Road, Bangor, LL57 2PW, Gwynedd, United Kingdom.

Insights

Incubator design significantly impacts neonatal imaging by causing beam attenuation, but its effect on radiation dose and image quality remains unclear. Further standardization is needed for optimal practice in imaging neonates.

Area of Science:

  • Medical Imaging
  • Neonatal Care
  • Radiology

Background:

  • Neonatal imaging is crucial for patient care.
  • Incubator design can influence imaging parameters.
  • Understanding these impacts is vital for optimizing diagnostic procedures.

Purpose of the Study:

  • To systematically review the impact of incubator design elements (canopy, mattress, mattress support) on neonatal imaging.
  • To assess effects on imaging technique, radiation dose, and image quality.

Main Methods:

  • Systematic literature review of multiple healthcare databases.
  • Inclusion of 7 eligible studies (6 experimental phantom-based, 1 retrospective analysis).

Main Results:

  • Incubator components caused beam attenuation ranging from 12% to 72%.
  • Image quality effects varied, with some studies showing reduced quality with incubator trays, while others found no significant difference.
  • Limited evidence exists on whether increased exposure factors are needed due to beam attenuation.

Conclusions:

  • Existing incubator designs cause significant beam attenuation, but its correlation with neonatal radiation dose and image quality is poorly understood.
  • Further research and standardization are necessary to ensure best practices in neonatal imaging.
Abstract