Safety assessment of spine MRI in deep brain stimulation patients

Alexandre Boutet1,2, Gavin J B Elias2, Robert Gramer2

  • 11Joint Department of Medical Imaging, University of Toronto.

Abstract

Insights

Spine MRI is safe for patients with deep brain stimulation (DBS) devices. Modified protocols with fewer slices allow safe imaging, avoiding adverse effects and ensuring device stability.

Area of Science:

  • Neurology
  • Radiology
  • Biomedical Engineering

Background:

  • Deep brain stimulation (DBS) devices present challenges for MRI due to safety concerns.
  • Existing guidelines often restrict routine MRI scans in patients with DBS.
  • Assessing MRI safety in DBS patients is crucial for comprehensive patient care.

Purpose of the Study:

  • To evaluate the safety of spine MRI in patients with implanted DBS devices.
  • To develop and validate a modified MRI protocol for DBS patients.
  • To determine potential risks such as device heating and adverse effects.

Main Methods:

  • Phantom model testing to assess DBS device heating during 1.5-T spine MRI.
  • Modification of routine clinical MRI sequences to achieve a temperature rise < 2°C.
  • Prospective validation in 9 DBS patients undergoing 67 spine MRI sequences.

Main Results:

  • Modified spine MRI protocols (2D sequences, reduced slices) achieved a temperature rise < 2°C.
  • No acute or long-term adverse effects were observed in DBS patients post-MRI.
  • Device impedances remained stable, and no acute brain MRI findings were noted.

Conclusions:

  • Clinically indicated spine MRI can be safely performed in patients with DBS devices.
  • Modified protocols, with fewer slices than vendor guidelines, enable safe imaging.
  • This approach may reduce the need for direct radiologist supervision during MRI acquisition.