Related Experiment Video
Updated: Dec 28, 2025

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
The in-ear region as a novel anatomical site for ECG signal detection: validation study on healthy volunteers
R De Lucia1, G Zucchelli2, V Barletta2
1Second Division of Cardiology, CardiacThoracic and Vascular Department, Azienda Ospedaliero Universitaria Pisana, Via Paradisa 2, 56100, Pisa, Italy. r.delucia.md@gmail.com.
Purpose:
Early detection of cardiac arrhythmias is a major opportunity for mobile health, as wearable devices nowadays available can detect single-lead electrocardiogram (ECG). The study aims to validate the in-ear region as a new anatomical site for ECG signal detection and looks towards designing innovative ECG wearable devices.
Methods:
We performed ECG using KardiaMobile device (AliveCor®) on 35 healthy volunteers. First, ECG was detected by standard modality using both hands. Then, ECG was detected using the left in-ear region instead of the right hand. All the recorded ECGs were analyzed by the device and by two cardiologists in blind testing.
Results:
We successfully collected 70 ECGs performed on 35 volunteers (male 54%, age 39.1 ± 10.7 years; BMI 22.9 ± 2.89 kg/m2) with no differences observed by KardiaMobile in ECG reports detected in the two different modalities. All the ECGs were reported as normal by the device and the two cardiologists. Moreover, linear regression analysis showed good correlation between the amplitude (mV) of P (r = 0.76; r2 = 0.57; p < 0.0001) and QRS waves (r = 0.81; r2 = 0.65; p < 0.0001), the intervals (ms) of PR (r = 0.91; r2 = 0.83; p < 0.0001; LOA - 0.60-0.41; CC = 0.91), QRS (r = 0.78; r2 = 0.61; p < 0.0001; LOA - 0.49-0.43; CC = 0.78), QT (r = 0.85; r2 = 0.71; p < 0.0001; LOA - 1.31-1.20; CC = 0.85), and heart rate (r = 0.94; r2 = 0.89; p < 0.0001; LOA - 7.82-7.76; CC = 0.94) detected in two different modalities.
Conclusion:
The in-ear region is a reliable novel anatomical site for ECG signal detection in normal healthy subjects. Further studies are needed to validate this new ECG detection modality also in case of cardiac arrhythmias and to support the development of new wearable devices.
More Related Videos
10:50Behavioral Determination of Stimulus Pair Discrimination of Auditory Acoustic and Electrical Stimuli Using a Classical Conditioning and Heart-rate Approach
Published on: June 6, 2012
04:59Author Spotlight: Exploring the Effects of Transauricular Vagus Nerve Stimulation
Published on: January 19, 2024
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...