Influence of field of view size on image quality: ultra-high-resolution CT vs. conventional high-resolution CT
Tomo Miyata1, Masahiro Yanagawa2, Akinori Hata3
1Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita City, Osaka, 565-0871, Japan. t-miyata@radiol.med.osaka-u.ac.jp.
Objectives:
This study was conducted in order to compare the effect of field of view (FOV) size on image quality between ultra-high-resolution CT (U-HRCT) and conventional high-resolution CT (HRCT).
Methods:
Eleven cadaveric lungs were scanned with U-HRCT and conventional HRCT and reconstructed with five FOVs (40, 80, 160, 240, and 320 mm). Three radiologists evaluated and scored the images. Three image evaluations were performed, comparing the image quality with the five FOVs with respect to the 160-mm FOV. The first evaluation was performed on conventional HRCT images, and the second evaluation on U-HRCT images. Images were scored on normal structure, abnormal findings, and overall image quality. The third evaluation was a comparison of the images obtained with conventional HRCT and U-HRCT, with scoring performed on overall image quality. Quantitative evaluation of noise was performed by setting ROIs.
Results:
In conventional HRCT, image quality was improved when the FOV was reduced to 160 mm. In U-HRCT, image quality, except for noise, improved when the FOV was reduced to 80 mm. In the third evaluation, overall image quality was improved in U-HRCT over conventional HRCT at all FOVs. Noise of U-HRCT increased with respect to conventional HRCT when the FOV was reduced from 160 to 40 mm. However, at 240- and 320-mm FOVs, the noise of U-HRCT and conventional HRCT showed no differences.
Conclusions:
In conventional HRCT, image quality did not improve when the FOV was reduced below 160 mm. However, in U-HRCT, image quality improved even when the FOV was reduced to 80 mm.
Key Points:
• Reducing the size of the field of view to 160 mm improves diagnostic imaging quality in high-resolution CT. • In ultra-high-resolution CT, improvements in image quality can be obtained by reducing the size of the field of view to 80 mm. • Ultra-high-resolution CT produces images of higher quality compared with conventional HRCT irrespective of the size of the field of view.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT