Paper-based electrodes modified with cobalt phthalocyanine colloid for the determination of hydrogen peroxide and
A Sánchez-Calvo1, A Costa-García, M C Blanco-López
1Instituto de Biotecnología de Asturias. Departamento de Química Física y Analítica, Universidad de Oviedo, 33006 Oviedo, Spain. cblanco@uniovi.es.
Abstract:
Cobalt(ii) phthalocyanine (CoPc) was suspended in aqueous medium and the colloidal system was used as catalyst for the electrochemical determination of hydrogen peroxide on paper-based electrodes modified with carbon nanomaterials. H2O2 was oxidised at 0.275 V vs. Ag pseudoreference electrode. This system was adapted to develop a glucose sensor with glucose oxidase immobilized on the cellulose electrode. CoPc suspended nanoparticles acted as nanoenzyme mimicking peroxidase activity and were combined with different carbon nanomaterials to form hybrids with optimised catalytic performance. GO-CoPc paper-based electrodes yielded the best results with a linear range of ∼12 μM to 49 mM for H2O2 and 0.1 mM to 1 mM for glucose. Glucose was determined in physiological serum and juice samples with recoveries of 93.3 and 94.2% respectively. CoPc could replace HRP for the catalytic sensing of H2O2, without the need to be dissolved. This material can be used in situ in a simple protocol with other nanomaterials for electrode modification. The sensor described has the advantage of easy preparation, using the catalyst in colloidal form, long term stability, and versatility to be adapted to other low cost and disposable enzymatic systems.
Related Concept Videos
Amperometry: Overview
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...


![Protein Film Infrared Electrochemistry Demonstrated for Study of H2 Oxidation by a [NiFe] Hydrogenase](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55858.jpg&w=3840&q=50)