Data transformations for statistical assessment of quantitative imaging biomarkers: Application to lung nodule
Qi Gong1, Qin Li1, Marios A Gavrielides1
1US Food and Drug Administration, CDRH/OSEL/DIDSR, Silver Spring, MD, USA.
Abstract:
Variance stabilization is an important step in the statistical assessment of quantitative imaging biomarkers. The objective of this study is to compare the Log and the Box-Cox transformations for variance stabilization in the context of assessing the performance of a particular quantitative imaging biomarker, the estimation of lung nodule volume from computed tomography images. First, a model is developed to generate and characterize repeated measurements typically observed in computed tomography lung nodule volume estimation. Given this model, we derive the parameter of the Box-Cox transformation that stabilizes the variance of the measurements across lung nodule volumes. Second, simulated, phantom, and clinical datasets are used to compare the Log and the Box-Cox transformations. Two metrics are used for quantifying the stability of the measurements across the transformed lung nodule volumes: the coefficient of variation for the standard deviation and the repeatability coefficient. The results for simulated, phantom, and clinical datasets show that the Box-Cox transformation generally had better variance stabilization performance compared to the Log transformation for lung nodule volume estimates from computed tomography scans.