Augmented reality (AR) in laparoscopic liver surgery shows increased registration errors with lower target accuracy. Reducing annotation errors and registration volume size is crucial for improving AR system reliability in surgical navigation.
Area of Science:
- Surgical Technology
- Medical Imaging
- Computer-Assisted Surgery
Background:
- Augmented reality (AR) systems are increasingly used in minimally invasive surgery.
- Accurate image-to-patient registration is critical for the effectiveness of AR in laparoscopic liver resection.
- Point-based registration (PBR) is a common method for achieving this registration.
Purpose of the Study:
- To evaluate the accuracy of point-based registration (PBR) for augmented reality (AR) applications in laparoscopic liver resection.
- To assess the impact of varying target sampling accuracy on registration performance.
Main Methods:
- The study utilized three scenarios with decreasing target accuracy: a machined phantom, a patient-specific phantom, and an in vivo study.
- Target Registration Error (TRE) and Fiducial Registration Error (FRE) were calculated.
- Image-to-patient registration was performed using five randomly selected positions.
Main Results:
- Mean TRE was 6.9 mm for the machined phantom, 7.9 mm for the patient-specific phantom, and 13.4 mm in the in vivo setting.
- AR demonstrated increased TRE and FRE as target sampling accuracy decreased.
- The size of the registration volume influenced the observed errors.
Conclusions:
- Augmented reality systems are sensitive to the accuracy of target sampling in point-based registration.
- Reducing errors in target annotation and the size of registration volumes can enhance the robustness of AR systems in surgery.