Acetaminophen interference with Nova StatStrip® Glucose Meter: case report with bench top confirmation.
James A Chenoweth1, Luke T Dang2, Guofeng Gao2
1Division of Medical Toxicology, Department of Emergency Medicine, University of California - Davis School of Medicine, Sacramento, CA, USA.
High acetaminophen levels can interfere with point-of-care glucose meters, including the Nova StatStrip®. This interference may occur even with devices previously thought to be resistant, impacting patient assessment during overdose cases.
Area of Science:
- Clinical Chemistry
- Toxicology
- Point-of-Care Testing
Background:
- Point-of-care glucose meters are crucial for assessing patients with altered mental status.
- Interference from common substances like acetaminophen is a known concern for glucose meters.
- The Nova StatStrip® glucose meter has been previously recognized for its resistance to interference.
Observation:
- A case report details a patient with a very high acetaminophen concentration (465 mg/L) presenting after an overdose.
- Attempts to measure glucose using the Nova StatStrip® glucose meter resulted in error messages across five devices.
- Laboratory analysis confirmed the high acetaminophen level and a glucose reading of 180 mg/dL.
Findings:
- Acetaminophen concentrations of 399 mg/L and above were found to interfere with the Nova StatStrip® glucose meter in laboratory spiking studies.
- This interference, particularly at very high acetaminophen levels, has not been previously documented for this specific device.
- The study highlights a potential limitation in the reliability of point-of-care glucose monitoring in severe acetaminophen toxicity.
Implications:
- Clinicians must be aware of potential Nova StatStrip® glucose meter interference in cases of significant acetaminophen overdose.
- Relying solely on point-of-care glucose readings in such scenarios could lead to misdiagnosis or delayed treatment.
- Confirmation of glucose levels via laboratory analysis may be necessary for patients with suspected or confirmed high acetaminophen concentrations.
More Related Videos
08:30Intraperitoneal Glucose Tolerance Test, Measurement of Lung Function, and Fixation of the Lung to Study the Impact of Obesity and Impaired Metabolism on Pulmonary Outcomes
Published on: March 15, 2018
12:08Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Related Concept Videos
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Hypoglycemia and Glucagon
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...