David Chong1,2, Sabrina Kayser1, Eniko Szakmar1,3
1Neonatal Intensive Care Unit, The Rosie Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK.
Background:
Pressure rise time (PRT), also known as slope time to the peak inflating pressure can be set on some modern neonatal ventilators. On other ventilators, PRT is determined by the set circuit flow. Changing slope time can affect mean airway pressure (MAP), oxygenation, and carbon dioxide elimination. Our aim was to investigate the effect of PRT on ventilator parameters and gas exchange during volume-guaranteed ventilation.
Methods:
In a crossover study, 12 infants weighing greater than 2 kg were ventilated using a Dräger Babylog VN500 ventilator with synchronized intermittent positive pressure ventilation with volume guarantee (SIPPV-VG) and pressure support ventilation with volume guarantee (PSV-VG). During both modes PRTs between 0.08 and 0.40 seconds were used in 15-minute epochs. Data from the ventilator and patient monitors were downloaded with 1- and 100-Hz sampling rate and analyzed using the Python computer language.
Results:
During PSV-VG, longer PRTs were associated with longer inspiratory time (P < .0001) and with lower peak inflating pressure (PIP; P = .003), but the MAP was similar. During SIPPV-VG the PIP was not significantly different; however, MAP was lower with longer PRT (P = .001). With a short PRT (0.08 seconds), the PIP was higher during PSV-VG than during SIPPV-VG (19.8 vs 16.5 mbar; P = .042). There were no significant differences in tidal volume delivery, respiratory rate, minute volume, oxygen saturations, or end-tidal CO2 with different PRTs in either mode.
Conclusions:
During SIPPV-VG or PSV-VG, using short or long PRTs affects some ventilation parameters but does not significantly change oxygenation or carbon dioxide elimination.
More Related Videos
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
08:59Method of Isolated Ex Vivo Lung Perfusion in a Rat Model: Lessons Learned from Developing a Rat EVLP Program
Published on: February 25, 2015
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Pulmonary Cycle: Exhalation