Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Comparing the postoperative refractive predictability of Pentacam HR and IOLMaster 500 after a multifocal intraocular
Newton Andrade Junior1, Wilson Takashi Hida2, André Marcio Vieira Messias3
1Hospital de Olhos Leiria de Abdrade, Fortaleza, CE, Brazil.
Purpose:
To compare the postoperative refractive predictability of IOLMaster 500 and Pentacam HR on the basis of keratometry and anterior chamber depth values in eyes with an indication for multifocal intraocular lens (IOL) implantation.
Methods:
This was a retrospective study conducted on 118 eyes treated with phacoemulsification and multifocal intraocular lens implantation. Only the eyes that achieved emmetropia in the dynamic refraction performed on postoperative day 30 were included. Haigis' formula was used in each case to calculate the intraocular lens power, and the intraocular lens with the target refraction closest to emmetropia was implanted. Four lens calculation scenarios were tested by combining keratometry and anterior chamber depth measurements obtained using the two devices.
Results:
IOLMaster 500 and Pentacam HR differed with regard to mean keratometry (D 0.07 ± 0.03 D; p=0.0065) and anterior chamber depth (D 0.08 ± 0.01 mm; p<0.001). In the analysis of covariance, the following differences were obtained using the Haigis' formula when confronted with the biometric values obtained by inserting keratometry and anterior chamber depth values, respectively: Penta/IOL x IOL/Penta (0.13 ± 0.03; p<0.0001); Penta/Penta × IOL/Penta (0.13 ± 0.03; p<0.0001); Penta/IOL × IOL/IOL (0.11 ± 0.03; p=0.001); Penta/Penta × IOL/IOL (0.11 ± 0.03; p=0.002); IOL/IOL × IOL/Penta (0.02 ± 0.03; p=0.865); and Penta/IOL × Penta/Penta (0.002 ± 0.03; p=0.99). The difference was smaller when measuring the anterior chamber depth using the IOLMaster 500, regardless of which device was used to measure keratometry.
Conclusions:
Pentacam HR significantly differed from IOLMaster 500 when calculating keratometry. As regards the anterior chamber depth, the two devices were equally accurate.