Related Experiment Video
Updated: Oct 14, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Errors in femoral anteversion, femoral offset, and vertical offset following robot-assisted total hip arthroplasty
Chih-Ming Hsieh1, Stephen M Howell1, Maury L Hull1,2,3
1Department of Biomedical Engineering, University of California, Davis, Sacramento, California, USA.
Abstract:
The objectives were to determine errors in femoral anteversion (FA), femoral offset (FO), and vertical offset (VO) with robot-assisted total hip arthroplasty (THA) and how consistently these errors are within clinically desirable limits of ±5° and ±5 mm. After preoperative planning, robot-assisted THAs were performed on twelve cadaveric specimens. The error between achieved and planned component placements was used to determine bias (mean error) and precision (SD of error). The percent of the population within clinically desirable limits was determined. Bias of 1.5° and 2.7 mm occurred for FA and VO, respectively. Precision was 1.2° for FA and better than 1.5 mm for FO and VO. The percent of population within clinically desirable limits was at least 95% for FA and at least 99% for FO. With limits of ±7 mm, at least 95% of the population was within these limits for VO. Robot-assisted THA may improve clinical outcomes.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
08:15Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023