Two-site evaluation of a new workflow for the detection of malignant cells on the Sysmex XN-1000 body fluid analyzer
Related Experiment Video
Updated: Dec 26, 2025

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Julien Favresse1, Lidvine Boland2, Marie Schellen2
1CHU UCL Namur, Namur Thrombosis and Hemostasis Center, Hematology Laboratory, Université catholique de Louvain, Yvoir, Belgium.
Introduction:
The presence of high fluorescent cells (HF-BF) on the Sysmex XN-1000 hematology analyzers has gained interest regarding the prediction of malignant cells in body fluids, but lacks sensitivity. We aimed to increase this sensitivity by combining HF-BF value, automated results, and clinical information.
Methods:
We evaluated a new workflow for the management of body fluids in the hematology laboratory, including the HF-BF criterion and clinical information. In two laboratories, 1623 serous fluids were retrospectively analyzed on the XN-1000 BF mode. All samples were morphologically screened for malignant cells. Optimal HF-BF cutoffs were determined to predict their presence. Thereafter, the added value of clinical information was evaluated. Other reflex testing rules (eosinophilic count >5% and presence of the WBC Abnormal Scattergram flag) were also used to refine our workflow.
Results:
Optimal HF-BF cutoffs in the two hematology centers were 108 and 45 cells/µL, yielding a sensitivity/specificity of 66.7/93.6% and 86.8/66.6% for malignant cell detection. When adding clinical information, sensitivity/specificity evolved to 100.0/68.9% and 100.0%/not determined. Of 104 samples containing malignant cells, 97 had positive clinical information; the remainder had a HF-BF > cutoff.
Conclusion:
Combining clinical information and HF-BF reached 100% sensitivity for malignant cell detection in body fluid analysis. Lack of robustness of the optimal HF-BF cutoff deserves the use of local cutoffs. Rapid automated results reporting from the XN-1000 BF mode are also feasible in clinical practice. Prospective evaluation of the workflow is needed before its implementation in clinical practice.