Related Experiment Video
Updated: Jun 17, 2026

One-day Workflow Scheme for Bacterial Pathogen Detection and Antimicrobial Resistance Testing from Blood Cultures
Published on: July 9, 2012
Evaluation of Allplex™ Entero-DR assay for detection of antimicrobial resistance determinants from bacterial cultures
María Fernanda Mojica1,2,3, Elsa De La Cadena4, Adriana Correa5,6
1Grupo de Resistencia Antimicrobiana y Epidemiología Hospitalaria, Universidad El Bosque, Ak. 9 #131a-02, Bogotá, DC, Colombia.
Objective:
To evaluate the sensitivity and specificity of the Allplex™ Entero-DR, a quantitative PCR-based method, for the detection of β-lactamase-encoding genes and vancomycin-resistance determinants in 156 previously characterized Gram-negative bacilli and Enterococcus spp. from bacterial cultures.
Result:
The method had 100% sensitivity and between 92 and 100% of specificity for identifying blaKPC, blaVIM, blaIMP, blaNDM, blaOXA-48-like, blaCTX-M and vanA. In nine isolates, unspecific amplifications were detected. The Ct of these false positives was above 33. The Ct of the correctly identified bla and van genes did not surpass 28 and 30, respectively. None of the clinical isolates included as negative controls yielded any amplification. Therefore, the Allplex™ Entero-DR assay is a highly accurate test for the detection of important antibiotic resistance determinants. With this assay, reliable results can be obtained within 3 h. However, according to our data, samples with Ct values greater than 33 should be considered with caution.
More Related Videos
12:14A Duplex Digital PCR Assay for Simultaneous Quantification of the Enterococcus spp. and the Human Fecal-associated HF183 Marker in Waters
Published on: March 9, 2016
08:58Isolation and Identification of Waterborne Antibiotic-Resistant Bacteria and Molecular Characterization of their Antibiotic Resistance Genes
Published on: March 3, 2023