Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Cyclic Fatigue Comparison of TruNatomy, Twisted File, and ProTaper Next Rotary Systems.
Abdullah Mahmoud Riyahi1, Amr Bashiri2, Khalid Alshahrani2
1Department of Restorative Dental Sciences, Division of Endodontics, College of Dentistry, King Saud University, P.O. Box 60169, Riyadh 11545, Saudi Arabia.
The new TruNatomy (TN) file demonstrated superior cyclic fatigue resistance compared to ProTaper Next (PTN) and Twisted Files (TF). This finding suggests TN offers an advantage in endodontic procedures, though further clinical evaluation is needed.
Area of Science:
- Endodontics
- Dental Materials Science
- Biomaterials Engineering
Background:
- Endodontic files undergo cyclic fatigue during root canal preparation.
- Evaluating cyclic fatigue resistance is crucial for assessing instrument durability and clinical performance.
- The TruNatomy (TN) system is a novel endodontic file that requires performance evaluation.
Purpose of the Study:
- To evaluate the cyclic fatigue resistance of the new TruNatomy (TN) file.
- To compare the cyclic fatigue resistance of TN with established systems: ProTaper Next (PTN) and Twisted Files (TF).
Main Methods:
- Forty-five endodontic files (15 each of PTN X2, TF, and TN prime file) were subjected to cyclic fatigue testing.
- Testing was performed in artificial stainless-steel canals with a 60-degree curvature and 5 mm radius at 300 rpm.
- The number of cycles to failure (NCF) was recorded for each file, and data were analyzed using ANOVA and Tukey's test.
Main Results:
- The TruNatomy (TN) file exhibited a significantly higher mean NCF (846.67 ± 37.16) compared to Twisted Files (TF) (521.67 ± 63.07) and ProTaper Next (PTN) (259 ± 37.2).
- Statistical analysis confirmed TN's superior cyclic fatigue resistance (p < 0.05) against both TF and PTN.
- All tested files had a length of 25 mm, with specific sizes and tapers noted for each group.
Conclusions:
- The TruNatomy (TN) file demonstrates superior cyclic fatigue resistance compared to ProTaper Next (PTN) and Twisted Files (TF).
- The enhanced cyclic fatigue resistance of TN may offer clinical advantages, potentially related to its design and tooth structure preservation.
- Further research is warranted to assess other properties and clinical efficacy of the TruNatomy system.