Using an Automated Hirschberg Test App to Evaluate Ocular Alignment
Published on: March 24, 2020
[The performance of visual photoscreening for Chinese preschool children aged 4 to 5 years]
1Department of Ophthalmology, the First Affiliated Hospital of Nanjing Medical University, Nanjing 210029, China.
Insights
Photoscreening is an effective tool for detecting refractive amblyopia risk factors (ARFs) in Chinese preschoolers. This study found photoscreening highly accurate when compared to cycloplegic retinoscopy, supporting its use in vision screening programs.
Area of Science:
- Ophthalmology
- Pediatric Optometry
- Vision Science
Background:
- Amblyopia remains a significant cause of preventable vision impairment in children.
- Early detection of refractive amblyopia risk factors (ARFs) is crucial for timely intervention.
- Photoscreening offers a potential method for mass vision screening in preschool populations.
Purpose of the Study:
- To evaluate the accuracy of photoscreening in identifying refractive amblyopia risk factors (ARFs) in 4- to 5-year-old Chinese children.
- To compare photoscreening results with cycloplegic retinoscopy (CR), considered the gold standard.
Main Methods:
- A cross-sectional study involving 1,986 preschool children in Nanjing, China.
- Photoscreening (Plusoptix A12C) was performed without cycloplegia, followed by CR for a subset of children.
- Statistical analysis included Wilcoxon signed rank test and Bland-Altman plots to assess agreement.
Main Results:
- Photoscreening demonstrated high testability (99.04%) and reliability (96.56% within three attempts).
- Bland-Altman analysis showed good agreement between photoscreening and CR measurements for sphere, cylinder, and spherical equivalent.
- The sensitivity and specificity of various referral criteria ranged from 53.36% to 97.87%, with an optimal astigmatism cut-off of 1.38 D identified.
Conclusions:
- Photoscreening is a viable and accurate tool for detecting refractive amblyopia risk factors in preschool children.
- The findings support the integration of photoscreening into routine preschool vision screening protocols.
- Further research can refine referral criteria for optimal sensitivity and specificity.
Abstract:
Objective: To assess the accuracy of photoscreening for detecting refractive amblyopia risk factors (ARFs) in Chinese preschool children aged 4 to 5 years. Methods: A cross-sectional study. Comprehensive ocular examinations were conducted for preschool children in Nanjing, China from September to December, 2016. Photoscreening (Plusoptix A12C) was applied for refractive screening without cycloplegia. Voluntary children and children suspected of eyes abnormalities received cycloplegic retinoscopy (CR). Results of photoscreening and CR were compared using Wilcoxon signed rank test, and Bland-Altman plot were used to assess the agreement between the photoscreener and CR. According to the updated preschool vision screening guidelines from American Association for Pediatric Ophthalmology and Strabismus (AAPOS) in 2013, CR was adopted for identifying children with ARFs, which was considered as a golden standard. Based on the golden standard, the accuracy of 5 sets of referral criteria (including sensitivity standard, Matta/Silbert standard, AAPOS2013 standard, Alaska Blind Child Discovery standard, specificity standard) for photoscreener were tested. Receiver operating characteristics curves were constructed applied to evaluate the quality of the photoscreener in refractive ARFs detection and to find probably the best cut-off points. Results: In total, 1 986 children [mean age, (4.57±0.29) years] received comprehensive examinations, including 1 084 boys and 902 girls. The test ability of photoscreening was 99.04% (1 967/1 986) in the preschool children, and 96.56%(1 827/1 892) of the children got a reliable result within three screening attempts. In 538 children who had data of CR, refractive error of one child exceeded the upper limit of the photoscreener value setting, which was directly categorized as hyperopia, so in the end, 537 children were included to analyze the comparison between the two tests. The measurement values of photoscreening were lower than those of CR in sphere, cylinder and spherical equivalent [(0.75 (0.50, 1.25) D vs. 1.25 (1.00, 1.75) D, Z=-10.36, P<0.01; -0.50 (-0.75, -0.25) D vs. -0.25 (-0.75, 0.00) D, Z=-11.10, P<0.01; 0.63 (0.38, 0.88) D vs. 1.00 (0.75, 1.50) D, Z=-13.33, P<0.01]. The 95% limit of agreement cover rates between the photoscreening and CR in sphere, cylinder and spherical equivalent was 96.28% (517/537), 95.34% (512/537) and 96.65% (519/537), respectively. Based on the golden standard, 47 (8.74%) children had refractive ARFs, and the range of sensitivity, specificity, Youden index, positive predictive values and negative predictive values for detecting refractive ARFs of the 5 common used referral criteria was from 63.83% to 97.87%, from 53.36% to 97.56%, from 0.51 to 0.80, from 16.73% to 74.51% and from 96.57% to 99.62%, respectively. Considering particular refractive ARFs on the basis of the receiver operating characteristic curves, the optimal cut-off point for astigmatism was set at 1.38 D. Conclusion: Photoscreening could be an applicable tool to detect refractive ARFs in preschool children. (Chin J Ophthalmol, 2020, 56: 189-196).