Related Experiment Video
Updated: Dec 26, 2025

Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
GeoGebra: A reliable and free software for measuring acetabular cup anteversion on digitalized plain radiographs
Vikas Bachhal1, Gaurav Saini2, Nipun Jindal3
1Department of Orthopaedics, PGIMER, Sector 12, Chandigarh, India.
Objective:
Although accurate measurement of cup anteversion in hip replacement requires CT scans, however, its routine application, especially during follow-up, remains economically and ethically unreasonable. Thus, several methods have been devised for making this measurement on plain radiographs. In recent years, several ways have been adopted using software on digital radiographs. We present one such method which uses open access mathematical software GeoGebra.
Methods:
Anteversion was measured on 72 radiographs (36 cemented; 36 uncemented) by three different observers using this software. One observer repeated measurements at three weeks interval.
Results:
The intraclass correlation coefficient for interobserver variability and intraobserver variability was 0.982 (0.973-0.989) and 0.986 (0.978-0.991) respectively. There was a significant difference in the reliability of the method for cemented and uncemented cups with higher reliability for cemented cups (p < 0.001).
Conclusion:
GeoGebra software can be used as a reliable alternative for measuring acetabular cup anteversion on good quality well centred digital radiographs of the pelvis.