Coronary Access After TAVR With a Self-Expanding Bioprosthesis: Insights From Computed Tomography
Mohammad Abdelghani1, Martin Landt2, Hussein Traboulsi2
1Heart Center, Segeberger Kliniken, Bad Segeberg, Germany; Department of Cardiology, Al-Azhar University, Cairo, Egypt; Cardiology Department, the Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands.
Objectives:
The authors sought to estimate possible interference of the Medtronic Evolut R/Pro transcatheter heart valve (THV) frame with coronary access using multislice computed tomography (MSCT) data.
Background:
Lower-risk patients undergoing transcatheter aortic valve replacement (TAVR) endure a high cumulative risk of coronary events, but coronary access can be challenging.
Methods:
In 101 patients who received an Evolut R/Pro THV, post-TAVR MSCT (performed at a median of 30 days after TAVR) was used to assess possible interference of the elements of the THV frame with coronary access.
Results:
The closest cell of the THV frame vertically aligned with the coronary ostium was located opposite the ostium in 58% and 63%, below the ostium in 22% and 30%, or above the ostium in 20% and 7% of left and right coronary arteries, respectively. The free sinus of Valsalva space between the THV frame and the coronary ostium was 0.45 ± 0.17 cm and 0.44 ± 0.17 cm for the left and right coronary arteries, respectively, and showed a stepwise decrease with decreasing THV size (p < 0.001). Bioprosthetic valve commissures were antianatomic (i.e., not aligned with native commissures) in 45 patients (47%), and the commissural post was overlapping a coronary ostium in 15 patients (16%). Two patients (2.0%) had a possible interference of the paravalvular sealing skirt with coronary access.
Conclusions:
Using post-TAVR MSCT data, the main mechanism of potential interference of Evolut R/Pro frame with coronary access was an antianatomic commissural post overlapping the coronary ostium.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...