Capillary Electrophoresis Separation of Monoclonal Antibody Isoforms Using a Neutral Capillary
Published on: January 16, 2017
In-depth analysis of monoclonal antibodies using microfluidic capillary electrophoresis and native mass spectrometry
Sara Carillo1, Craig Jakes2, Jonathan Bones2
1National Institute for Bioprocessing Research and Training, Fosters Avenue, Mount Merrion, Blackrock, Co. Dublin, Ireland.
Charge variant profiling of therapeutic proteins is required by the International Council for Harmonisation guidelines and is traditionally performed by capillary electrophoresis or ion exchange chromatography. Recently, improvements in the hyphenation of capillary electrophoresis with mass spectrometry and the introduction of mass spectrometry compatible background electrolytes has allowed the implementation of native mass spectrometric determination of the charge variant profile obtained from the electrophoretic separation. The low flow operation of the microfluidic electrophoretic platform significantly boosts mass spectrometric sensitivity and increases the dynamic range, even when using sample amounts as low as 1 ng in capillary. In the current study, rituximab, trastuzumab and bevacizumab drug products were analysed using the ZipChip microfluidic CE-ESI-MS platform that facilitated confident identification of proteoforms with an average mass accuracy of <15 ppm. Up to 52 proteoforms were identified for trastuzumab drug product, while rituximab sample revealed the presence of fragments and sialylated N-glycans. Overall, the CE-ESI-MS platform proved to be a fast and robust tool for therapeutic protein charge variant profiling and facilitated efficient coupling with native mass spectrometry for the generation of highly informative characterisation data.
Charge variant profiling of therapeutic proteins is required by the International Council for Harmonisation guidelines and is traditionally performed by capillary electrophoresis or ion exchange chromatography. Recently, improvements in the hyphenation of capillary electrophoresis with mass spectrometry and the introduction of mass spectrometry compatible background electrolytes has allowed the implementation of native mass spectrometric determination of the charge variant profile obtained from the electrophoretic separation. The low flow operation of the microfluidic electrophoretic platform significantly boosts mass spectrometric sensitivity and increases the dynamic range, even when using sample amounts as low as 1 ng in capillary. In the current study, rituximab, trastuzumab and bevacizumab drug products were analysed using the ZipChip microfluidic CE-ESI-MS platform that facilitated confident identification of proteoforms with an average mass accuracy of <15 ppm. Up to 52 proteoforms were identified for trastuzumab drug product, while rituximab sample revealed the presence of fragments and sialylated N-glycans. Overall, the CE-ESI-MS platform proved to be a fast and robust tool for therapeutic protein charge variant profiling and facilitated efficient coupling with native mass spectrometry for the generation of highly informative characterisation data.
More Related Videos
04:24Author Spotlight: Advancing Biotherapeutic Mass Calculation by Introducing mAbScale, a Python-Based Desktop Application
Published on: June 16, 2023
10:50Purification and Analytics of a Monoclonal Antibody from Chinese Hamster Ovary Cells Using an Automated Microbioreactor System
Published on: May 1, 2019
Related Concept Videos
MALDI-TOF Mass Spectrometry
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...