Design and practical evaluation of a shielded application system for intravenously administered radionuclide
Christian Kühnel1, Thomas Winkens1, Tobias Niksch1
1Clinic of Nuclear Medicine, Jena University Hospital, Jena, Germany.
Aim:
Intravenous radionuclide therapies are gaining importance. With the increased frequency of these therapies, safe application procedures in combination with effective radiation protection are needed. We present a shielded system which can be universally used for the application of liquid radiopharmaceuticals.
Materials And Methods:
The application system consists of a cuboidal box made of lead with stainless steel coating, an inner layer of polymethyl methacrylate (PMMA), a disposable line/bottle system, and a medical infusion pump. No proprietary disposable parts are used. The system was tested and validated ex-vivo, and evaluated for Lu-177 peptide receptor radionuclide and for radioligand therapies (PRRT, RLT).
Results:
Construction of the application system was performed in accordance with the physical characteristics of the used isotopes. 10 validation procedures with 1 GBq Tc-99m-pertechnetate confirmed functionality. 38 PRRT and 13 RLT procedures were performed successfully and completely. At least 98 % of the prescribed activities were infused. No leakage of radioactivity occurred. Dose rate measurements showed that the radiation in the box is shielded completely, and that exposure arises only from the infusion line outside the box and from the patient.
Conclusion:
The presented application system for intravenous radionuclide therapies is feasible, safe, and cost-efficient. It has been shown that Lu-177-PRRT and -RLT can be performed without complications while ensuring nearly complete infusion of the prescribed radiopharmaceutical dose. Radiation exposure of the applying physician and other staff is low. The construction of the shielding box ensures complete shielding of all radionuclides currently used for radiomolecular therapies.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...