Calibration of the LiF - thermoluminescent detectors used for personal dose equivalent Hp(3) assessment
1University of Lodz, Facility of Physics and Applied Informatics, Department of Nuclear Physics and Radiation Safety, Pomorska 149/153, 90-236 Lodz, Poland.
Aims:
The issue of exposure of eye lenses of employees exposed to ionizing radiation is an interesting topic not only from the point of view of deterministic effects related to the occurrence of cataracts, but also dosimetric aspects, in particular the calibration of detectors in units enabling the assessment of eye lens exposure or personal dose equivalent Hp(3). The paper presents the idea of calibrating thermoluminescent detectors designed for the Hp(3) values measurement of gamma radiation, which the source is the process of annihilation of positrons emitted by the deoxyglucose marker - 18F radionuclide.
Methods:
The method was based on the value of air kerma Ka to Hp(3) conversion coefficients (Hp(3,0°)/Ka) developed as part of the ORAMED project. High-sensitivity thermoluminescent detectors (MCP-N) produced in Poland were used in the measurements. During the exposure of the detectors, a 137Cs gamma radiation source (irradiator 137Cs/60Co) and a 20cm diameter cylinder filled with water were used.
Results & Conclusions:
The value of conversion coefficient Hp(3,0°)/Ka for energy 511 keV is 1.31Sv/Gy and the calibration factor is (3.46±0.03)·10-4 mSv/N (N - number of counts). Verification of the value of the obtained coefficient carried out using a cylinder with a diameter of 20cm showed a difference of less than 2% in relation to the value obtained by the method described in this paper.