The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Difficult airway management practice changes after introduction of the GlideScope videolaryngoscope: A retrospective
Alexander Avidan1, Yoel Shapira, Avital Cohen
1From the Department of Anesthesiology, Critical Care and Pain Medicine, Faculty of Medicine, Hadassah Medical Center, Hebrew University of Jerusalem (AA, YS, AC, CW) and Intensive Care Unit, Shaarei Zedek Medical Center, Jerusalem, Israel (PDL).
Background:
Introduction of the GlideScope videolaryngoscope caused a change in use of other devices for difficult airway management.
Objective:
The influence of the GlideScope videolaryngoscope on changes in the indications for and the frequency of use of flexible fibreoptic-assisted intubation and other difficult airway management techniques.
Design:
Retrospective cohort study.
Setting:
Tertiary care referral centre.
Methods:
Two periods of equal length (647 days each) before and after introducing the GlideScope were compared. Information about patients who were intubated using nondirect laryngoscopic techniques were analysed. Data were retrieved from the anaesthesia and hospital information management systems.
Results:
Difficult airway management techniques were used in 235/8306 (2.8%) patients before and in 480/8517 (5.6%) (P < 0.0001) patients after the introduction of the GlideScope. There was an overall 44.4% reduction in use of flexible fibreoptic bronchoscopy after GlideScope introduction [before 149/8306 (1.8%); after 85/8517 (1.0%), P < 0.0001]. The GlideScope replaced flexible fibreoptic bronchoscopy in most cases with expected and unexpected difficult intubation. In patients with limited mouth opening, flexible fibreoptic bronchoscopy was still mostly the first choice after the introduction of the GlideScope. There was a 70% reduction in the use of other difficult intubation techniques after the introduction of the GlideScope [before 84/8306 (1.0%); after 22/8517 (0.3%), P < 0.0001)].
Conclusion:
The GlideScope videolaryngoscope replaced flexible fibreoptic bronchoscopy for most patients with expected and unexpected difficult intubation. In the case of limited mouth opening, flexible fibreoptic bronchoscopy was still the first choice after the introduction of the GlideScope. The reduced use of flexible fibreoptic bronchoscopy raises concerns that residents may not be adequately trained in this essential airway management technique. GlideScope use was disproportionately greater than the reduction in the use of flexible fibreoptic bronchoscopy and other difficult intubation techniques. This may be attributed to resident teaching and use in patients with low-to-moderate suspicion of difficult intubation.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Procedures III: Video Capsule Endoscopy
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required