Testing a phantom eye under various signal-to-noise ratio conditions using eleven different OCT devices
Tuomas Heikka1,2, Giovanni Ometto3,4,2, Giovanni Montesano3,4
1Department of Ophthalmology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
Abstract:
We compared eleven OCT devices in their ability to quantify retinal layer thicknesses under different signal-strength conditions, using a commercially available phantom eye. We analyzed a medium-intensity 50 µm layer in an identical manner for all devices, using the provided log-scale images and a reconstructed linear-scale tissue reflectivity metric. Thickness measurements were highly comparable when the data were analyzed in an identical manner. With optimal signal strength, the thickness of the 50 µm layer was overestimated by a mean of 4.3 µm in the log-scale images and of 2.7 µm in the linear-scale images.