Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Clinical Reliability of point-of-care tests to support community based acute ambulatory care
J Y Verbakel1, C Richardson2, T Elias3
1MD, PhD, Nuffield Department of Primary Care Health Sciences, University of Oxford, Oxford, United Kingdom.
Objective:
To ensure clinicians can rely on point-of-care testing results, we assessed agreement between point-of-care tests for creatinine, urea, sodium, potassium, calcium, Hb, INR, CRP and subsequent corresponding laboratory tests.
Participants:
Community-dwelling adults referred to a community-based acute ambulatory care unit.
Interventions:
The Abbott i-STATTM (Hb, clinical chemistry, INR) and the AfinionTM Analyser (CRP) and corresponding laboratory analyses.
Outcomes:
Agreement (Bland-Altman) and bias (Passing-Bablok regression).
Results:
Among 462 adults we found an absolute mean difference between point-of-care and central laboratory analyses of 6.4g/L (95%LOA -7.9 to +20.6) for haemoglobin, -0.5mmol/L (95%LOA -4.5 to +3.5) for sodium, 0.2mmol/L (95%LOA -0.6 to +0.9) for potassium, 0.0mmol/L (95%LOA -0.3 to +0.3) for calcium, 9.0 micromol/L (95%LOA -18.5 to +36.4) for creatinine, 0.0mmol/L (95%LOA -2.7 to +2.6) for urea, -0.2 (95%LOA -2.4 to +2.0) for INR, -5.0 mg/L (95%LOA -24.4 to +14.4) for CRP.
Conclusions:
There was acceptable agreement and bias for these analytes, except for haemoglobin and creatinine.