Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Screening of obstructive sleep apnea in patients who snore using a patch-type device with electrocardiogram and
Ying-Shuo Hsu1,2, Tien-Yu Chen3,4, Dean Wu5,6,7
1Department of Otolaryngology, Shin Kong Wu Ho-Su Memorial Hospital, Taipei, Taiwan.
Chest wall motion (CEI) and cyclic variation of heart rate (CVHR) detected using a single-lead ECG patch accurately identify obstructive sleep apnea (OSA). CEI demonstrated superior diagnostic accuracy compared to CVHR for moderate-to-severe OSA.
Area of Science:
- Cardiology
- Pulmonology
- Biomedical Engineering
Background:
- Obstructive sleep apnea (OSA) diagnosis is challenging due to a lack of comfortable and accessible screening tools.
- Current diagnostic methods often require extensive equipment and specialized facilities.
Purpose of the Study:
- To evaluate the diagnostic accuracy of chest wall motion and cyclic variation of heart rate (CVHR) for detecting OSA.
- To compare the efficacy of these two metrics using a single-lead electrocardiogram (ECG) patch with a 3-axis accelerometer.
Main Methods:
- 119 patients underwent polysomnography with a single-lead ECG patch.
- Chest wall motion was measured using the chest effort index (CEI) via a 3-axis accelerometer.
- CVHR signals were collected from the ECG patch, and both indices were compared for diagnosing moderate-to-severe OSA (apnea-hypopnea index ≥ 15) using area under the curve (AUC).
Main Results:
- CVHR showed 52.9% sensitivity and 94.1% specificity (AUC: 0.76) for moderate-to-severe OSA.
- CEI demonstrated 80% sensitivity and 79.4% specificity (AUC: 0.87), significantly outperforming CVHR (ΔAUC: 0.11, P = .032).
- Combining CEI and CVHR improved discrimination for severe OSA (AUC = 0.90).
Conclusions:
- Both CEI and CVHR, derived from a patch-type device with ECG and accelerometer, can effectively detect moderate-to-severe OSA.
- Incorporating CEI enhances the detection of sleep apnea using a single-lead ECG with a 3-axis accelerometer.
More Related Videos
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
04:33Author Spotlight: Unveiling the Connection Between Sleep Disorders and Cognitive Symptoms in Depression
Published on: April 26, 2024
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...