Related Experiment Video
Updated: Dec 24, 2025

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Post hoc closure of large bore vascular access using the MANTA closure device
Ravi K Sharma1, Marie-France Poulin1, Hector Tamez-Aguilar1
1Cardiovascular Institute, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts, USA.
Abstract:
Successful closure of large bore vascular access (≥12 Fr) is facilitated by the MANTA vascular closure device; however, it requires a critical first step of measuring "skin to arterial wall" depth of the access site using the depth locator before dilating the vessel above 8 Fr. It may be challenging at times to acquire the deployment depth at the onset of the procedure due to case urgency, delayed closure, or when large bore access is obtained at a different institution. We discuss a novel technique of measuring the deployment depth and successful delayed closure of large bore arterial access using the MANTA closure device.