Biomechanical analysis of occipitocervical stabilization techniques: emphasis on integrity of osseous structures at
Bryan W Cunningham1, Kyle B Mueller2, Kenneth P Mullinix1
11Musculoskeletal Education Center, Department of Orthopaedic Surgery, MedStar Union Memorial Hospital, Baltimore, Maryland.
Objective:
The objective of the current study was to quantify and compare the multidirectional flexibility properties of occipital anchor fixation with conventional methods of occipitocervical screw fixation using nondestructive and destructive investigative methods.
Methods:
Fourteen cadaveric occipitocervical specimens (Oc-T2) were randomized to reconstruction with occipital anchors or an occipital plate and screws. Using a 6-degree-of-freedom spine simulator with moments of ± 2.0 Nm, initial multidirectional flexibility analysis of the intact and reconstructed conditions was performed followed by fatigue loading of 25,000 cycles of flexion-extension (x-axis, ± 2.0 Nm), 15,000 cycles of lateral bending (z-axis, ± 2.0 Nm), and 10,000 cycles of axial rotation (y-axis, ± 2.0 Nm). Fluoroscopic images of the implantation sites were obtained before and after fatigue testing and placed on an x-y coordinate system to quantify positional stability of the anchors and screws used for reconstruction and effect, if any, of the fatigue component. Destructive testing included an anterior flexural load to construct failure. Quantification of implant, occipitocervical, and atlantoaxial junction range of motion is reported as absolute values, and peak flexural failure moment in Newton-meters (Nm).
Results:
Absolute value comparisons between the intact condition and 2 reconstruction groups demonstrated significant reductions in segmental flexion-extension, lateral bending, and axial rotation motion at the Oc-C1 and C1-2 junctions (p < 0.05). The average bone mineral density at the midline keel (1.422 g/cm3) was significantly higher compared with the lateral occipital region at 0.671 g/cm3 (p < 0.05). There were no significant differences between the occipital anchor and plate treatments in terms of angular rotation (degrees; p = 0.150) or x-axis displacement (mm; p = 0.572), but there was a statistically significant difference in y-axis displacement (p = 0.031) based on quantitative analysis of the pre- and postfatigue fluoroscopic images (p > 0.05). Under destructive anterior flexural loading, the occipital anchor group failed at 90 ± 31 Nm, and the occipital plate group failed at 79 ± 25 Nm (p > 0.05).
Conclusions:
Both reconstructions reduced flexion-extension, lateral bending, and axial rotation at the occipitocervical and atlantoaxial junctions, as expected. Flexural load to failure did not differ significantly between the 2 treatment groups despite occipital anchors using a compression-fit mechanism to provide fixation in less dense bone. These data suggest that an occipital anchor technique serves as a biomechanically viable clinical alternative to occipital plate fixation.
More Related Videos
Related Concept Videos
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
Cranial Bones: Superior and Posterior View
The frontal bone is the single bone that forms the forehead. At its anterior midline, between the eyebrows, there is a slight depression called the glabella. The frontal bone also forms the supraorbital margin of the orbit. Near the middle of this margin is the supraorbital foramen, the opening that provides passage for a sensory nerve to the forehead. The frontal bone is thickened just above each supraorbital margin,...