Dual-energy CT imaging of orbits during episcleral brachytherapy with Ru-106 plaques: A phantom study on its
Kostas Perisinakis1, Efstathios T Detorakis2, Antonis Tzedakis3
1University of Crete, Medical School, Department of Medical Physics, 71003 Heraklion, Crete, Greece.
Purpose:
To investigate the potential of dual energy CT (DECT) to suppress metal artifacts and accurately depict episcleral brachytherapy Ru-106 plaques after surgical placement.
Methods:
An anthropomorphic phantom simulating the adult head after surgical placement of a Ru-106 plaque was employed. Nine DECT acquisition protocols for orbital imaging were applied. Monochromatic 140 keV images were generated using iterative reconstruction and an available metal artifact reduction algorithm. Generated image datasets were graded by four observers regarding the ability to accurate demarcate the Ru-106 plaque. Objective image quality and visual grading analysis (VGA) was performed to compare different acquisition protocols. The DECT imaging protocol which allowed accurate plaque demarcation at minimum exposure was identified. The eye-lens dose from orbital DECT, with and without the use of radioprotective bismuth eye-shields, was determined using Monte Carlo methods.
Results:
All DECT acquisition protocols were judged to allow clear demarcation of the plaque borders despite some moderate streaking/shading artifacts. The differences between mean observers' VGA scores for the 9 DECT imaging protocols were not statistically significant (p > 0.05). The eye-lens dose from the proposed low-exposure DECT protocol was found to be 20.1 and 22.8 mGy for the treated and the healthy eye, respectively. Bismuth shielding was found to accomplish >40% reduction in eye-lens dose without inducing shielding-related artifacts that obscure plaque delineation.
Conclusions:
DECT imaging of orbits after Ru-106 plaque positioning for ocular brachytherapy was found to allow artifact-free delineation of plaque margins at relatively low patient exposure, providing the potential for post-surgery plaque position verification.