Sound comparison of seven TMS coils at matched stimulation strength
Lari M Koponen1, Stefan M Goetz2, Debara L Tucci3
1Department of Psychiatry & Behavioral Sciences, Duke University, Durham, NC, 27710, USA.
Background:
Accurate data on the sound emitted by transcranial magnetic stimulation (TMS) coils is lacking.
Methods:
We recorded the sound waveforms of seven coils with high bandwidth. We estimated the neural stimulation strength by measuring the induced electric field and applying a strength-duration model to account for different waveforms.
Results:
Across coils, at maximum stimulator output and 25 cm distance, the sound pressure level (SPL) was 98-125 dB(Z) per pulse and 76-98 dB(A) for a 20 Hz pulse train. At 5 cm distance, these values were estimated to increase to 112-139 dB(Z) and 90-112 dB(A), respectively.
Conclusions:
The coils' airborne sound can exceed some exposure limits for TMS subjects and, in some cases, for operators. These findings are consistent with the current TMS safety guidelines that recommend the use of hearing protection.
More Related Videos
Related Concept Videos
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...
Motor Unit Stimulation
The latent period of contraction marks the onset of excitation-contraction coupling, when the action potential propagates across the sarcolemma, preparing the muscle fibers for contraction. As the fibers enter the contraction phase, the...
Magnetic Field Due to Two Straight Wires
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...