HoloLens-Based Vascular Localization System: Precision Evaluation Study With a Three-Dimensional Printed Model.
Taoran Jiang1, Dewang Yu2, Yuqi Wang2
1Department of Plastic and Reconstructive Surgery, Shanghai 9th People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Journal of Medical Internet Research
|April 18, 2020
Summary
Augmented reality with HoloLens precisely localizes perforators for surgery. This system, tested on a 3D model, showed minimal navigation errors, suggesting its potential for safer surgical procedures.
Area of Science:
- Medical Technology
- Surgical Navigation
- Augmented Reality
Background:
- Vascular localization is critical for successful perforator flap transfer.
- Augmented reality (AR) systems, like the HoloLens headset, can integrate computed tomographic angiography data with the surgical field.
- This AR integration allows surgeons to visualize subsurface vasculature, enhancing perforator identification.
Purpose of the Study:
- To evaluate the precision of the HoloLens-based augmented reality system for vascular localization.
- To determine the system's accuracy as a key performance indicator for surgical navigation.
Main Methods:
- A HoloLens AR system was used to localize vessels on a 3D printed model in a simulated operating room.
- Navigation error was calculated by measuring the distance between corresponding points on virtual and physical vascular maps.
- Testing occurred under various conditions, including different visual angles, light intensities, and motion states.
Main Results:
- The HoloLens system demonstrated a mean navigation error between 1.35 mm and 3.18 mm, falling within clinically acceptable limits.
- No significant differences in error were observed across varying visual angles, light intensities, or static/motion states.
- A notable finding was that the system performed with greater accuracy in the absence of light.
Conclusions:
- The HoloLens-based system exhibits high precision for perforator localization, supporting its use in surgical navigation.
- The system has the potential to improve the safety and convenience of perforator flap transfer procedures.
- The study recommends utilizing the HoloLens surgical navigation system without light for optimal performance.